Features
· Microcomputer control functions
· Color liquid crystal display and touch-sensitive panel
· User name is shown on screen, linked to user ID code
· Powerful customization functions
· Low table height easier rotor handling
· Real-time control (RTC) simplifies timer settings
· Operating parameters are automatically computed and displayed, based on selected RCF
· Important functions are shown on screen
· The highest acceleration and RCF for its class 100,000 rpm, and RCF to 803,000 x g
· Drive unit warranted for 10 years
· Powerful imbalance protection
· Rotor life is automatically managed
· Rotor life is automatically extended
· Safety and environmental impact reduction
· 0°C is maintained by electronic cooling
· Energy savings
· Exceptionally low noise 53 dB or less
· Based on EU requirements for mechanical and electrical safety and electromagnetic compatibility
· Brake control system for reverting motor electric power to the power source side
· Ecoproduct
Specifications
Model
|
CP100WX
|
CP90WX
|
CP80WX
|
Max. speed (rpm)
|
100,000
|
90,000
|
80,000
|
Max.RCF (xg)
|
803,000
(P100AT2 rotor)
|
700,000
(P90AT rotor)
|
615,000
(P80AT rotor)
|
Speed control accuracy
|
±10 rpm (1,000 rpm to maximum speed)
|
Set speed
|
1,000 rpm to maximum speed in increments of 100 rpm
|
Displaying and setting RCF
|
Applicable (RCF max. or RCF average)
|
ω2T run
|
Applicable
|
Drive unit warranty
|
10 years
|
Rotor temperature control / display accuracy
|
±0.5ºC (set temperature is from 0ºC to 40ºC)
|
Cooling method
|
Thermo-module cooling (CFC/HCFC-free)
|
Operation panel
|
. Color liquid crystal display (256 color) and keyboard
. Touch-sensitive panel (run screen only)
|
Rotor life management
|
RLM rotor: automatic management,
Rotor with optical adaptor: management by registration system
|
Operation results management system /
calculation & simulation system
|
Optional software (CD-ROM, OS: Windows® XP Professional / 2000 Professional)
This option requires a PC and RS232C cross cable on the market.
|
Vacuum system
|
Oil-rotary vacuum pump with moisture removal function and oil diffusion pump.
|
HEPA filter
|
Option
|
Operational noise (dB(A))
|
53
|
Heat radiation into the room
|
1 kW or less
|
Dimensions (mm)
|
790 (W) x 690 (D) x 1,000 (H), depth with safety cover: 890, height to the table: 850
|
Floor area (m2)
|
0.81 (900 x 900 mm)
|
Weight (kg)
|
400
|
Power requirements
|
Single phase, 50/60 Hz,
AC200/208/220 V ±10%, 20 A max. (normally 8 A)
AC 230/240 V ±10%, 16 A max. (normally 7A)
|
Conformity to EMC * standards
|
The CP-WX Series ultracentrifuges conform to the following EMC standards:
EN61326, EN61000-3-2, and EN61000-3-3
|